【摘要】心臟除顫器作為一種急救類醫療設備,廣泛應用于各醫療場所。其各項計量性能指標的準確性不僅影響急救的效果,還與病人的身體健康安全息息相關。本文針對該類設備相應的參數進行測量不確定度的分析與評定,以提高校準結果的準確性,為相關計量檢定人員提供參考。
【關鍵詞】心臟除顫器;不確定度;示值誤差
【DOI編碼】10.3969/j.issn.1674-4977.2024.05.077
Evaluation of Measurement Uncertainty for Defibrillators
KUANG Shimin
(Hainan Provincial Inspection Institute,Haikou 570206 China)
Abstract: As emergency medical devices, defibrillators are widely used in various medical settings. The accuracy of their various metrological performance indicators not only affects the effectiveness of emergency treatment but is also closely related to the health and safety of patients. This article analyzes and evaluates the measurement uncertainty of the corresponding parameters for such devices to enhance the accuracy of calibration results, providing a reference for metrological verification personnel.
Keywords: defibrillator; uncertainty; indication error
根據校準規范,除顫分析儀釋放能量最大允許誤差為±2 J或測量值的±5%,兩者取大值。因此,在心臟除顫器0~400 J的測量范圍內應分兩部分來進行釋放能量的不確定度分析與評定。
1.1在(0,40] J測量范圍內釋放能量的測量不確定度評定

1.1.1測量不確定度的分量
1)由心臟除顫器釋放能量測得值重復性引入的不確定度分量
在除顫分析儀(0,40] J的測量范圍內,將被校儀器釋放能量擋選擇為40 J,在該能量點測量10次釋放能量,測量結果分別為39.8、39.6、39.7、39.2、39.5、39.9、39.6、39.3、39.7、39.5 J。計算得單次測量結果的實驗標準偏差為s( E ) = 0.22J。因校準過程中所需測量值為單次測量的結果,故所引入標準不確定度分量為:





【參考文獻】
[1]心臟除顫器校準規范:JJF 1149—2014[S].
[2]測量不確定度評定與表示:JJF 1059.1—2012[S].
[3]古麗孜拉.心臟除顫器和心臟除顫監護儀測量不確定度評定[J].計量與測試技術,2015,42(8):78-80.
【作者簡介】
鄺仕敏,男,1988年出生,工程師,學士,研究方向為醫學計量檢定。