999精品在线视频,手机成人午夜在线视频,久久不卡国产精品无码,中日无码在线观看,成人av手机在线观看,日韩精品亚洲一区中文字幕,亚洲av无码人妻,四虎国产在线观看 ?

A novel multifunctional electronic calibration kit integrated by MEMS SPDT switches*

2021-11-23 07:32:40ShanShanWang王姍姍QianNanWu吳倩楠YueShengGao高躍升JianGangYu余建剛QianLongCao曹釬龍LuLuHan韓路路andMengWeiLi李孟委
Chinese Physics B 2021年11期

Shan-Shan Wang(王姍姍) Qian-Nan Wu(吳倩楠) Yue-Sheng Gao(高躍升)Jian-Gang Yu(余建剛) Qian-Long Cao(曹釬龍)Lu-Lu Han(韓路路) and Meng-Wei Li(李孟委)

1School of Instrument and Electronics,North University of China,Taiyuan 030051,China

2Nantong Institute of Intelligent Opto-Mechatronics,North University of China,Nantong 226000,China

3Center for Microsystem Intergration,North University of China,Taiyuan 030051,China

4Academy for Advanced Interdisciplinary Research,North University of China,Taiyuan 030051,China

5School of Science,North University of China,Taiyuan 030051,China

Keywords: microelectromechanical system (MEMS), electronic calibration kit, single-pole double-throw(SPDT)switch,short-open-load-through(SOLT)calibration

1. Introduction

With the rapid development of electronics, medicine,industries, automotive, and aerospace technology, the requirements for miniaturization and high reliability of systems are increasing, and these requirements can be fundamentally achieved by high-degree integration of systems.The microelectromechanical system(MEMS)technology has gained popularity in the electronic information field, such as in the fabrication of integrated circuits, to produce miniaturized, intelligent, and integrated MEMS devices. At present, devices based on MEMS technology have gradually moved from the laboratory to practical application level, and they greatly contribute to the development of fields such as industry and agriculture, information, environment, bioengineering, medical treatment, space technology, and national defense.[1-4]The MEMS-based devic12-es such as radio-frequency microelectromechanical system (RF MEMS)switches,[5-7]attenuators,[8,9]filters,[10,11]and phase shifters[12-15]have been developed. However,the network analyzer, microwave test cable, and load used in the RF performance testing process all encounter problems of large loss,instability,and non-repeatability,and these problems directly result in the low accuracy of MEMS device testing.[16-18]Therefore, the test cable, load, test instrument, and other systems need calibrating prior to the MEMS device testing in order to minimize the system errors generated during the test,thereby improving the measurement accuracy of the MEMS device.[19]

Currently, most widely used calibration kits for MEMS devices are mechanical[20,21]and electronic[22]calibration kits, based on the coaxial calibration principle. Mechanical calibration kits are easy to operate and have high dependency on operator skills, whereas electronic calibration kits use automated operation, which is fast and less affected by operator skills. However,for the calibration of non-coaxial MEMS devices, both calibration kits encounter the problems of low matching degree and high error rate. In response to this,a through-reflect-line (TRL) calibration kit for S-parameter characterization of membrane circuits for the WR-03 waveguide band (220 GHz-325 GHz) was developed at Chalmers University of Technology.[23]Testing results verified that the calibrator has good response characteristics and reliability,but low sensitivity. A printed circuit board (PCB) was used to design and produce an electronic calibration kit suitable for power amplifiers(PA)at Malaysian University of Science and Technology.[24]Although the measurement accuracy of this calibration system was approximately 10%higher than that of traditional calibration kits,its size was at least six times larger than that of a sub-miniature A(SMA)interface connector. A non-contact detection calibration kit for 220 GHz-325 GHz was designed and produced at Ohio State University,[25]which effectively improves the mismatch caused by the wear of the probe tip. However,its operation is complicated and requires human operator judgment, which greatly affects the repeatability of the calibration kit.

To address the above-mentioned problems associated with existing MEMS device calibration methods,in this study a multifunctional electronic calibration kit is proposed and designed by using cascaded MEMS-based SPDT switches. The SOLT calibration states can be completed simultaneously,and the metallization of probe tip and errors introduced by human operation can be reduced by using the MEMS electronic calibration. It can calibrate a vector network analyzer,microwave probe station,and test cable in a frequency range of 0.1 GHz-20 GHz with high precision,and the overall size of this novel proposed calibrator kit is only 6 mm×2.8 mm×0.8 mm.

2. Theory

At present, the most effective and direct testing method of RF parameters of MEMS devices in the millimeter wave band is to use a vector network analyzer, a microwave probe station, a test cable, and a test method for connecting the device under test. The loss introduced by the cable and the probe will directly result in inaccurate test results. The commonly used two-port calibration methods are SOLT and TRL(straight-reflect-reflection line).[26]The SOLT calibration kit is easy to operate,and can provide excellent accuracy and repeatability, so it is more commonly used. Figure 1(a) shows the calibration kit for a probe station. The ground wire-signal wire-ground wire(GSG)probe is directly calibrated on a calibration sheet on which the short-circuit, open-circuit, load,and through(thru)units are arranged and distributed as shown in Fig. 1(b). In the dual-port calibration process performed by the vector network analyzer,the GSG probes connected to the port of the mobile network analyzer calibrate short-circuit,open-circuit, and load at ports 1 and 2; then, the two probes are placed on the thru calibration unit simultaneously. The four calibration units are immersed in sequence seven times to measure the S-parameters; subsequently, 12 errors are calculated according to the SOLT calibration formula. Thus, the calibration of the vector network and the probe station are completed.[27,28]The functions of the calibration ports are as follows.

Short calibration: The signal achieves an ideal short circuit on the calibration kits(total reflection|Γ|=1)and undergoes total reflection,yielding return loss S11close to 0.

Open calibration: The signal achieves an ideal open circuit(total reflection|Γ|=1)on the calibration kit and undergoes total reflection,yielding return loss S11close to 0.

Load calibration: The signal realizes an accurate broadband impedance matching with the system impedance on the calibration kit.

Fig.1. (a)Thru calibration of probe,and(b)calibration on-chip calibration kits.

2.1. Design of electronic calibration kit

The RF MEMS devices have characteristics of small size,low loss,and easy integration,and have important applications in the development of miniature,intelligent,light-weight,and low-power-consumption communication systems.[29]Compared with traditional electronic switches,RF MEMS switches have high frequency,small size,low power consumption,and high-degree integration,and they are used in many electronic devices. In this study is proposed an RF MEMS switch that can be introduced into an electronic calibration kit to realize a small-volume,low-loss,convenient,and multifunctional integrated electronic calibration kit as shown in Fig.2. The calibration kit consists of a standard short-circuit port Z0,a standard open port Z1,and a standard load port Z2.

Fig.2. Overall scheme of calibration kit.

The signal from port 1 passes through switches S1and S3and enters into Z0, providing the short-circuit calibration of port 1. The signal from port 2 passes through switches S9and S7to reach Z0, providing the short-circuit calibration of port 2. The signal from port 1 passes through switches S1and S4to arrive at Z1, providing the open calibration of port 1. The signal from port 2 enters into switches S9and S8and reaches Z1, realizing the open calibration of port 2. The signal from port 1 enters into switches S2and S6and reaches Z2,realizing the load calibration of port 1. The signal from port 2 enters into switches S10and S12and reaches Z2, providing the load calibration of port 2. Finally, the signal enters into switches S2,S5,S11,and S10from port 1 to port 2,which can be directly calibrated to complete the network signal dual-port calibration. This calibration kit is thus simple and convenient.

2.2. Design and simulation of RF MEMS SPDT switch

The RF MEMS SPDT switch is a three-port structure in which signals are input from one port and output to either of the other two ports selectively. The signal is input from the input port and is divided into two channels for transmission by the T-type power divider,and each channel has an RF MEMS switch. By controlling the opening or closing of the RF MEMS switch of each channel,the selected output of signal is realized.

Figure 3 is a schematic diagram of the typical series contact RF MEMS SPDT switch. When sufficient driving voltage is applied between the upper plate and lower plate of the RF MEMS switch,the upper electrode will move vertically downward under the action of electrostatic force, and the signal is turned on and the switch is closed as shown on port 2. When the driving voltage between the upper plate and lower plate is removed, the upper electrode is restored to the original position under the action of the restoring force, the signal is disconnected,and the switch is in the disconnected state as shown in port 3.

Fig.3. Schematic diagram of RF MEMS SPDT switch.

To meet the requirements for the application of the standard calibration kit, the ADS software is used to calculate the basic physical dimensions of the substrate and coplanar waveguide at a matching impedance of 50 Ω. In this study is adopted a series contact-type MEMS SPDT switches in the electrostatic drive mode, and the upper electrode has a simple straight-plate structure. It is composed of a glass substrate(dielectric constant is 4.6, thickness is 500 μm), a coplanar waveguide reflection line(20μm/120μm/20μm),a straightplate-type top electrode,an actuation electrode, air bridges,a contact point, and a driver. The electrodes and other kit are shown in Fig.4.

Fig.4. Structure of SPDT switch.

Fig.5. (a)Insertion loss and(b)isolation of SPDT switch.

The results of the simulation conducted by using the ANSOFT HFSS software are shown in Fig.5.When the input signal enters into port 1,the insertion loss is 0.36 dB at 20 GHz through port 2,with isolation 20 dB at 20 GHz,and the insertion loss is 0.37 dB at 20 GHz through port 3, with isolation 21.83 dB at 20 GHz. The results indicate that the two ports of the designed RF MEMS SPDT switch have good insertion loss and isolation, and excellent consistency with each other as well.

2.3. Design of load resistance

The load resistance is an important factor and should be considered in calibration under a load state. In this study, a tantalum nitride thin-film resistor with self-passivation, high corrosion resistance, and high stability is selected in the design.[30]The resistance of the thin-film resistor is related to its size and geometry and can be calculated from

whereR0is the square resistance of the thin-film resistor(100 Ω/□tantalum nitride resistor is selected in this study),lis the length of the thin-film resistor (50 μm), andwis the width of the thin-film resistor(100μm).

The ANSOFT HFSS software is used to simulate a 50-Ω load resistance,and the simulation results are shown in Fig.6.At 0.1 GHz-20 GHz, the return loss is within 10.95 dB-11.07 dB. The resistor has good stability and can be used as a load standard for calibration.

Fig.6. Simulation results of load resistance.

2.4. Design of electronic calibration kit based on RF SPDT switch

Figure 7 shows the structure of the entire electronic calibration kit, which is composed of symmetrically cascading five RF MEMS SPDT switches. For the vector network analyzer to perform single-port calibration,three calibrations are required. First, port 1 probe of the vector network analyzer port contacts port 1,and switches 1-4 are all open at the same time. the calibration kit is in a short-circuit state, and singleended short-circuit calibration is performed. Control switches 1 and 3 are both open. Next,switch 5 is switched off and the calibration kit is in the open state,then single-ended open circuit calibration is performed. Control switches 1, 3, 5, and 6 are all open, the calibration kit is in load resistance state,and 50-Ω load calibration is performed. For the single-ended calibration by the vector network analyzer, the calibration is performed seven times. The switch status is shown in Table 1.The vector network analyzer ports CH 1 and CH 2 contact port 1 and port 2 of the calibrator,respectively. Then,the two ports are calibrated in the states of shorted circuit,open circuit,load,and thru by closing the control switches. This process requires only two probe contacts, thus reducing the effect of human errors on the calibration accuracy.

The switch status of each calibration state is shown in Table 1,where 1 indicates that the switch is on and 0 represents that the switch is off.

The ANSOFT HFSS software is used to simulate the RF performance of the electronic calibration kit based on the MEMS SPDT switch,and the simulation results are shown in Fig. 8. Figure 8(a) shows the return loss of the calibration kit in the short-circuit state. Figure 8(b) indicates the return loss of the calibration kit in the open-circuit state. Figure 8(c)displays the return loss of the calibration kit in the load-circuit state,and figure 8(d)exhibits the insertion loss in the thru state.

Fig.7. Structure of electronic calibration kit based on RF SPDT switch.

Fig.8. Simulation results of calibration kit: return loss in(a)short state,(b)open state,and(c)load state,and(d)insertion loss in thru state.

Table 1. Corresponding switch mode of calibration state.

The simulation results indicate that for a frequency range of 0.1 GHz-20 GHz, the electronic calibrator can be reused and needs only two steps for one-time calibration. Further,the return loss is less than 0.18 dB in the short-circuit state is less than 0.18 dB and less than 0.035 dB in the open-circuit state.This suggests a lower reflection error. When the load resistance is 50 Ω, the echo is within 11.5 dB-13.5 dB, showing good stability. In the thru state, the insertion loss less than 0.27 dB has a lower reflection error. This calibration kit exhibits an excellent performance during dual-port calibration.

3. Fabrication and measurement of SPDT switch

In order to verify the simulation results of this novel multifunctional electronic calibration kit, the core device of the RF MEMS SPDT switch is fabricated by micro-nano surface technology,[31]and the fabrication process consists of nine steps and is shown in Fig.9.

Fig.9. Fabrication processes of SPDT switch.

Step (a) The surface of the BF33 glass sheet is cleaned to remove various inorganic impurities and organic contaminants.

Step(b) A 400-nm-thick silicon nitride film layer is deposited by PECVD at 350°C,and dry etching is performed to obtain bumps contacting the ohmic series switch.

Step (c) A 500-nm-thick aluminum film layer is deposited by magnetron sputtering and the wet etching is performed by using concentrated phosphoric acid to obtain the driving electrode and PAD of SPDT.

Step(d) A 300-nm-thick silicon nitride film layer is deposited again by PECVD,servins as an isolation layer,which is used to prevent metal-to-metal contact from diffusing and also to prevent aluminum metal from being destroyed by subsequent processes.

Step (e) A 50-nm-thick titanium adhesion film layer is deposited on the isolation layer, followed by a 150-nm-thick gold layer. The purpose of Ti/Au metal layer is to serve as a seed layer for electroplating coplanar waveguide(CPW).

Step (f) A 2-μm-thick Au layer is selectively deposited as CPW, and the Au layer also covers the exposed contact points of the series ohmic switches to provide low contact resistance.

Step(g) A Ti/Au seed layer is then removed by wet etching. Finally, the isolation nitride spacer is defined by photolithography and dry etching to expose the underlying PAD.The over-etch time of dry etching is determined by considering the removal of Si3N4capping and exposure of the underlying aluminium layer. The sacrificial layer for the definition of air gap is formed by 4-μm thick polyimide(PI).In order to solve the problem of PI step coverage, the spin-coated PI wafer is horizontally placed in a covered petri dish, and then the petri dish is placed in an oven at a constant temperature of 50°C for 4 h. The temperature and time parameters are selected based on the fact that the liquid PI has the fastest flow rate at 50°C and will become solid after 4 h,so 50°C and 4 h are chosen.

Step (h) The anchor vias in the sacrificial layer are obtained by a mask and wet etching,and followed closely by the alkaline developer dissolving the upper photoresist and continuing to etch the underlying PI.

Step(i) The beams of the switches and the air bridges are formed when a 2-μm-thick gold layer is electroplated. Then the seed layer is removed by wet etching.

Step(j) The sacrificial layer is released. At this point the manufacturing process of the switch is completed. The scanning electron microscopy(SEM)image of the prepared SPDT switch is shown in Fig.10.

Fig.10. SEM photograph of SPDT switch.

Fig.11. (a)Measured insertion loss response of SPDT switch,and(b)measured isolation response of SPDT switch.

The test results are shown in Fig. 11, which shows that the insertion loss of the MEMS SPDT switch is less than 1.0 dB and the isolation is more than 18 dB, which satisfies the design specifications and calibration requirements. The calibration sheet used for the probe can be used only once,

whereas the MEMS SPDT switch prepared in this study can be used 106times; the insertion loss changes after multiple tests has become small and the repeatability is good.The measurements and the simulations accord well each other, which makes the process and technology of designing multifunctional electronic calibration kit more confident.

4. Conclusions and perspectives

The electronic calibration kit in four calibration states are proposed that is composed of the five cascaded MEMS SPDT switches and one load resistance. In the working frequency band of 0.1 GHz-20 GHz,the return loss of the calibration kit is less than 0.18 dB in the shorted circuit state and less than 0.035 dB in the open circuit state, the insertion loss is less than 0.27 dB in the thru state. The electronic calibrator can be reused 106times and needs only two steps for one-time calibration. The calibration method has addressed the problems of the traditional calibration with a single function and many cumbersome calibration steps. Therefore,the multifunctional calibration kit can provide a convenient calibration method for the microwave test systems,such as vector network analyzers,RF probe stations,and millimeter wave test cables.

Acknowledgment

We thank the Key Laboratory of Instrumentation Science and Dynamic Measurement for their technique support.

主站蜘蛛池模板: 天堂中文在线资源| 伊人久久大线影院首页| 国产精品免费福利久久播放| 亚瑟天堂久久一区二区影院| 国产亚洲美日韩AV中文字幕无码成人 | 国产精品午夜福利麻豆| 国产不卡一级毛片视频| 国产白浆视频| 国产亚洲精品自在久久不卡| 久久精品只有这里有| 国产成人亚洲欧美激情| 成人综合在线观看| 四虎影视库国产精品一区| 婷婷六月在线| 四虎永久在线视频| 天天综合网色| 精品人妻一区无码视频| 久久综合伊人77777| 天堂va亚洲va欧美va国产| 波多野结衣无码AV在线| 啊嗯不日本网站| 国产成人凹凸视频在线| 69视频国产| 在线欧美一区| 嫩草在线视频| 爱色欧美亚洲综合图区| 精品三级网站| 久久久久人妻一区精品| 欧美午夜一区| 国产黄色免费看| 一本久道久久综合多人| 久久精品91麻豆| 精品伊人久久久久7777人| 国产欧美日韩视频怡春院| 中文字幕第4页| 亚洲经典在线中文字幕| 亚洲精品日产AⅤ| 日韩在线观看网站| 日本三级欧美三级| 伊人91视频| 夜夜高潮夜夜爽国产伦精品| 大香网伊人久久综合网2020| 18黑白丝水手服自慰喷水网站| 国产一级妓女av网站| 91亚洲精选| 日本久久网站| 精品欧美视频| 亚洲最新地址| 国产又色又爽又黄| 中文字幕人成人乱码亚洲电影| 一级毛片免费高清视频| 国产无码在线调教| 性视频一区| 中国毛片网| 一级一级特黄女人精品毛片| 五月综合色婷婷| 一区二区三区高清视频国产女人| 亚洲啪啪网| 国产精品一区二区国产主播| 国产主播在线观看| 精品福利一区二区免费视频| 蜜臀AV在线播放| 天堂网亚洲系列亚洲系列| 永久免费无码成人网站| 欧美在线综合视频| 亚洲精品免费网站| 伊人色天堂| 成人精品区| 久久夜色精品国产嚕嚕亚洲av| 欧美在线网| 久久久久亚洲AV成人人电影软件 | 国产精品视频999| 欧美成人午夜在线全部免费| 欧美三級片黃色三級片黃色1| 亚洲AⅤ综合在线欧美一区 | 国产成人精品高清在线| 亚洲无码精彩视频在线观看| 国产免费久久精品44| 亚洲人人视频| 波多野结衣的av一区二区三区| 国产欧美日韩综合在线第一| 欧美人与牲动交a欧美精品|