【摘要】電導率儀是測量液體溶液電導率的主要儀器設備,電導率儀電導池常數對測量數據的準確性有著重大影響。本文依據JJG 376—2007《電導率儀》,對電導率儀電導池常數進行校準,并給出了應用的測量不確定度。
【關鍵詞】電導率儀;電導池常數;測量不確定度
【DOI編碼】10.3969/j.issn.1674-4977.2024.05.076
Evaluation of Uncertainty Measurement on Conductivity Meter Conductance Cell Constant
CHENG Shuozhen, ZHAO Sihai, CHEN Yun, LYU Ting, CHEN Yuan
(Hubei Institute of Measurement and Testing Technology, Wuhan 430200, China)
Abstract: Conductivity meter is the main equipment for measuring the conductivity of liquid solution. The conductivity cell constant of the conductivity meter has a great influence on the accuracy of the measurement data. In this paper, based on JJG 376—2007 Electrolytic Conductivity Meters, the uncertainty generated in the measurement process of conductivity meter conductance cell constant is evaluated.
Keywords: conductivity meter; conductivity cell constant; uncertainty measurement
電導率儀是通過電化學測量方法測定電解質溶液電導率的儀器,在水質分析、污水處理、環境檢測和工農業生產等領域有著廣泛的應用[1-5]。在電導率測量過程中,選擇正確合適的電導率儀電極對于獲得高精度測量結果至關重要。通常,電導率儀電極在出廠時都會被標注相應的電導池常數,然而隨著電極長時間在惡劣的測量環境中使用,電極表面的敏感材料很容易被污染或者失效,導致電導池常數發生了變化,使得測量數據出現偏差。因此,在電導率儀的使用過程中,除了需要對電極進行日常維護,還必須對電極相應的電導池常數進行校準,從而保證測量數據的準確性。本文依據JJG 376—2007《電導率儀》[6],對電導率儀電導池常數進行校準,并給出了應用的測量不確定度。
1.1測量依據
JJG 376—2007《電導率儀》,JJF 1059.1—2012《測量不確定度評定與表示》。
1.2測量環境條件
溫度19~22℃,相對濕度40%~50%,附近無顯著電磁干擾,無顯著振動源存在。
1.3測量標準





電導率儀是測量液體溶液電導率的主要儀器設備,電導率儀電導池常數對測量數據的準確性有著重大影響。本文依據JJG 376—2007《電導率儀》,對電導率儀電導池常數進行校準,并給出了應用的測量不確定度。從評定過程和不確定度分量一覽表中可以看出,標準溶液標準值引入的不確定度對電導池常數的不確定度影響很大,測量重復性引入的不確定度影響偏小,主要原因是電導率儀和電極的性能都很好,其次測量過程中環境溫度和水浴溫度能夠得到很好的控制,使測量重復性數據小。溫度對標準溶液的電導率影響很大,因而在電導率儀電導池常數的校準過程中一定要注意溫度的控制,預留充足的時間平衡標準溶液的溫度。
【參考文獻】
[1]金東春,焦闊,王旭東,等.基于電導率控制的生活污水回用至脫硫系統研究[J].工業水處理,2018,38(2):106-108.
[2]郭照河,伊利軍,隗玉霞.電導率在水質分析質量控制中的應用[J].水資源保護,2004(1):21.
[3]大氣降水電導率的測定方法:GB/T 13580.3—1992[S].
[4]許潔茹,凌仕剛,王少飛,等.鋰電池研究中的電導率測試分析方法[J].儲能科學與技術,2018,7(5):926-955.
[5]馬萬征,趙蘆,戴珊珊,等.不同電導率對溫室黃瓜生長發育的研究[J].蘭州文理學院學報(自然科學版),2018,32(4)45-48.
[6]電導率儀檢定規程:JJG 376—2007[S].
【作者簡介】
程碩楨,女,1990年出生,工程師,博士,研究方向為化學計量與檢定。
(編輯:李加鵬)