中圖分類號(hào):TB92;TB133文獻(xiàn)標(biāo)志碼:A
Evaluation and calibration of the measurement characteristics of white light interference systems
CAI Xiaoyu,WU Junjie,WEI Jiasi (Key Laboratory of Bioanalysis and Metrology for State Market Regulation, Shanghai Institute of Measurement and Testing Technology, Shanghai , China)
Abstract: The measurement system based on the white light interference principle (such as white light interference microscope) has the technical advantages of sub-nanometer longitudinal resolution, non-contact, high efficiency of field scanning,etc.,and is often used to measure the surface morphology and structure of products on the micro/nano scale. The analysis, evaluation, and calibration of the measurement characteristics of white light interferometry instruments is a complex and practical work, which is the basis of the research and application of this kind of instruments.Based on the principle of white light interference and the definition and regulation of relevant national standards, the measurement characteristics of the white light interference measurement system and the quantities influencing the system are analyzed theoretically. On this basis, the measurement characteristics of WLIS for metrological application are presented, i.e., measuring range and indicating error. Then, the theoretical analysis of the above measurement characteristics and their measurement and calibration methods are discussed. The three-axis measurement deviation of the system is calibrated by the standard template of step height and line interval, and the 3D measurement traceability chain based on the white light interference system is constructed.
Keywords: white light interference; measurement characteristics; calibration; traceability
引言
基于白光干涉原理的測(cè)量系統(tǒng),如白光干涉顯微鏡,是利用白光干涉信號(hào)定位被測(cè)樣品的高度,從而表征測(cè)量被測(cè)樣品微納尺度的表面形貌[1],尤其在垂直方向具有亞納米級(jí)的分辨力。白光干涉測(cè)量系統(tǒng)(white light interference system,WLIS)是基于光學(xué)信號(hào)的、顯微視場(chǎng)的掃描方式,相較于探針式的測(cè)量系統(tǒng)具有非接觸、高效率的優(yōu)點(diǎn)。此外,WLIS在大氣環(huán)境下使用,對(duì)環(huán)境限制小[2-3],無(wú)需特別制樣,應(yīng)用場(chǎng)景廣泛,如集成電路、MEMS、精密加工件、微納材料等的表面表征測(cè)量[4]。
當(dāng)WLIS作為定量測(cè)量的儀器時(shí),其計(jì)量特性反映了儀器的測(cè)量功能和性能,對(duì)其計(jì)量特性的把握是使用該儀器的前提。在生產(chǎn)制造領(lǐng)域,只有確定了計(jì)量?jī)x器的計(jì)量特性,才能進(jìn)一步確定其合適的應(yīng)用場(chǎng)合(如作為社會(huì)公共計(jì)量標(biāo)準(zhǔn)、工作計(jì)量標(biāo)準(zhǔn)等),發(fā)揮其量值溯源的作用;在科學(xué)研究領(lǐng)域,計(jì)量特性也是評(píng)判實(shí)驗(yàn)結(jié)果準(zhǔn)確性、一致性和可靠性的依據(jù)—一實(shí)驗(yàn)測(cè)得結(jié)果與所用的WLIS的計(jì)量特性密切相關(guān)。綜上所述,研究、評(píng)價(jià)與校準(zhǔn)WLIS的計(jì)量特性是WLIS研究或應(yīng)用的重要基礎(chǔ)工作。
WLIS的工作原理
WLIS通過(guò)一系列光學(xué)組件實(shí)現(xiàn)參考光束與測(cè)量光束的干涉,其典型光路如圖1所示。白光光源發(fā)出的光束經(jīng)過(guò)偏振分光鏡或半透半反膜分為兩束相干光,其中一束到參考反射鏡后被反射到顯微干涉物鏡,另一束則透射到被測(cè)樣品表面,經(jīng)樣品表面反射后回到顯微干涉物鏡,于是,這兩束光在物鏡視場(chǎng)中會(huì)合發(fā)生干涉[5-7]

樣品表面的光強(qiáng)信號(hào)由白光光譜中不同頻率光波發(fā)生干涉后合成,場(chǎng)掃描中某一點(diǎn)的光強(qiáng)信號(hào)滿足式(1)[8-9]

式中: I(z) 為高度位置為 z 時(shí)的光強(qiáng); I1 為參考光的光強(qiáng); I2 為測(cè)量光的光強(qiáng); Re[R(τd)] 為光源自相干函數(shù)實(shí)部, τd 與當(dāng)前位置和參考距離h0 有關(guān)。根據(jù)式(1),可以通過(guò)對(duì)光強(qiáng)信號(hào)的分析解構(gòu)得出被測(cè)表面的相對(duì)高度信息。
工作時(shí),WLIS沿 Z 向?qū)Ρ粶y(cè)樣品進(jìn)行掃描,覆蓋樣品表面最低與最高的位置,對(duì)每個(gè)像素點(diǎn)的干涉信號(hào)進(jìn)行計(jì)算分析,確定每個(gè)點(diǎn)光強(qiáng)峰值相應(yīng)的 Z 向掃描高度,解構(gòu)整個(gè)掃描場(chǎng)內(nèi)的相對(duì)表面形貌。因此,白光干涉測(cè)量系統(tǒng)除了光學(xué)組件形成干涉,還包括運(yùn)動(dòng)掃描部件執(zhí)行定位并掃描、圖像傳感部件獲取表面圖像、下位控制器獲得同步的圖像與位置信息、上位機(jī)處理干涉信號(hào),解構(gòu)三維形貌。圖2是典型的WLIS組成框圖,其中,物鏡移相器提供納米級(jí)分辨率的 z 軸掃描步進(jìn),樣品臺(tái)則具備 x 、 y 、 z 三軸運(yùn)動(dòng)功能,實(shí)現(xiàn)定位與對(duì)焦,有些樣品臺(tái)還具備傾斜運(yùn)動(dòng)功能,用于調(diào)整樣品與物鏡中參考鏡的平行度,獲得更易分辨的干涉條紋。

2 WLIS的測(cè)量影響量分析
首先,研究分析WLIS的主要測(cè)量影響量及其來(lái)源,這些測(cè)量影響量將最終影響具體的計(jì)量特性,間接地反映在測(cè)量結(jié)果中[10]
根據(jù)系統(tǒng)的組成,系統(tǒng)影響量的來(lái)源主要有光學(xué)成像組件、圖像傳感器、控制采集軟件、分析校準(zhǔn)軟件(含校準(zhǔn)標(biāo)樣)、測(cè)量運(yùn)動(dòng)平臺(tái)及儀器整體。根據(jù)各部分的原理、特性或功能進(jìn)一步確定其可能的影響量[1],如表1所示。此外,被測(cè)樣品的表面WLI特性及其與測(cè)量系統(tǒng)的互相作用,也會(huì)影響系統(tǒng)的計(jì)量特性。
這些影響量有的屬于硬件特性,如來(lái)自光學(xué)器件和運(yùn)動(dòng)機(jī)構(gòu)的影響量;有的屬于軟件算法,主要是控制和分析軟件算法中的變量與方法;有的涉及掃描參數(shù)設(shè)定,如采樣間距、Z方向掃描步距等單次測(cè)量可設(shè)參數(shù);有的來(lái)自環(huán)境因素,如振動(dòng)。通過(guò)提高硬件技術(shù)參數(shù)、優(yōu)化硬件設(shè)計(jì)、增加誤差修正等方法來(lái)減小來(lái)自硬件的影響量,通過(guò)優(yōu)化軟件算法盡可能減小來(lái)自軟件算法的影響量,通過(guò)改善環(huán)境、隔離環(huán)境噪聲的方法減小環(huán)境因素的影響。然而,作為一個(gè)整體的測(cè)量系統(tǒng),各影響量或各組成之間互相關(guān)聯(lián),在優(yōu)化過(guò)程中影響量可能互相制約、此消彼長(zhǎng),如掃描速度、 Z 方向步距、掃描時(shí)間、測(cè)量算法與運(yùn)動(dòng)機(jī)構(gòu)的遲滯之間的矛盾,這也是大部分儀器研發(fā)調(diào)試和使用中均存在的問(wèn)題。

3 WLIS的計(jì)量特性分析、評(píng)價(jià)與 校準(zhǔn)
3.1 WLIS的計(jì)量特性分析
結(jié)合上述測(cè)量影響量,從應(yīng)用與量值溯源的角度,分析確定WLIS的計(jì)量特性。
首先,推薦性國(guó)標(biāo)[1]給出了一些WLIS的計(jì)量特性參數(shù),對(duì)其定義、測(cè)量結(jié)果的實(shí)際影響等加以剖析與闡述。
(1)放大系數(shù):從響應(yīng)曲線得到的線性回歸直線的斜率。放大系數(shù)這一參數(shù)引用了光學(xué)顯微鏡領(lǐng)域的技術(shù)術(shù)語(yǔ)和概念,結(jié)合WLIS的測(cè)量功能,放大系數(shù)可分為 x 軸、 y 軸和 z 軸三軸的放大系數(shù) αx 、 ay 、 az 。該組參數(shù)一定程度上反映了WLIS的靈敏度。
(2)線性偏差:實(shí)際響應(yīng)曲線與其線性回歸直線的最大偏差。線性偏差也被稱為非線性,區(qū)域線性偏差程度決定了WLIS在對(duì)應(yīng)各測(cè)量區(qū)域(點(diǎn))的測(cè)量準(zhǔn)確度。
(3)殘余平面:區(qū)域基準(zhǔn)的平面度。作為區(qū)域法表面形貌測(cè)量手段,殘余平面決定了表面形貌的測(cè)量準(zhǔn)確性。這一參數(shù)以
為位置點(diǎn)序列,只需評(píng)價(jià) z 方向數(shù)據(jù)的質(zhì)量。
(4)測(cè)量噪聲:儀器正常工作狀態(tài)下附加于輸出信號(hào)的噪聲。根據(jù)WLIS測(cè)量原理與模式,通常,測(cè)量時(shí) x ! y 方向不進(jìn)行掃描,故不計(jì)x 軸、 y 軸方向的噪聲,與殘余平面類似,測(cè)量噪聲也只評(píng)價(jià) z 方向數(shù)據(jù)。
(5)橫向周期限:儀器高度( z 方向)響應(yīng)降到 50% 時(shí)正弦輪廓的空間周期。橫向周期限與光學(xué)系統(tǒng)的數(shù)值孔徑相關(guān),反映了系統(tǒng)的橫向分辨能力。
(2號(hào) (6)x-y 垂直度: x 軸與 y 軸之間的角度和90° 的偏差。當(dāng)WLIS進(jìn)行測(cè)量區(qū)域拼接時(shí),WLIS的 x 軸或 y 軸需要進(jìn)行掃描運(yùn)動(dòng),此時(shí),x-y 垂直度即為 x 軸運(yùn)動(dòng)方向與 y 軸運(yùn)動(dòng)方向夾角與 90° 的偏差,這一參數(shù)也稱為 x ! y 軸正交性,決定了成像的保真率和拼接的精準(zhǔn)性。
從實(shí)際校準(zhǔn)的角度,其他應(yīng)該予以校準(zhǔn)和確認(rèn)的計(jì)量特性參數(shù),即WLIS在量值傳遞時(shí)需要考慮的一些參數(shù),還包括WLIS的縱向(Z向)測(cè)量范圍、橫向 (x/y 向)測(cè)量范圍、 x/y/z 軸測(cè)量偏差、縱向測(cè)量分辨力等。
3.2WLIS主要計(jì)量特性的評(píng)價(jià)與校準(zhǔn)
3.2.1 面向應(yīng)用的計(jì)量特性
源的WLIS參數(shù)主要有 x 、y、 z 三軸各自的測(cè)量范圍和示值誤差(或測(cè)量偏差),而測(cè)量分辨力、噪聲、軸間串?dāng)_等都作為測(cè)量不確定度分量體現(xiàn)在測(cè)量結(jié)果中[12]。測(cè)量范圍和示值誤差,尤其是示值誤差,是WLIS實(shí)施量值溯源的關(guān)鍵,其測(cè)量結(jié)果的不確定度會(huì)作為測(cè)量?jī)x器引入的不確定度分量傳遞到下一級(jí)被測(cè)對(duì)象中。
3.2.2 測(cè)量范圍
WLIS的縱向測(cè)量范圍的上限主要取決于測(cè)量運(yùn)動(dòng)平臺(tái)的縱向運(yùn)動(dòng)范圍、測(cè)量軟件系統(tǒng)的處理量、測(cè)頭工作距離等因素中的最小值。一般情況下,基于WLI測(cè)頭的物鏡工作距離限制了WLIS測(cè)量范圍上限,如表2所示。

縱向測(cè)量范圍的下限則受限于系統(tǒng)的縱向分辨力,具體的評(píng)價(jià)方法如下。首先重復(fù)掃描臺(tái)階高度標(biāo)準(zhǔn)樣板,獲得兩組測(cè)量數(shù)據(jù) [x,y,z] [x,y,z′] 。將兩組數(shù)據(jù)對(duì)應(yīng)像素點(diǎn)的高度值相減得到差值表面 [x,y,Δz] ,評(píng)價(jià)差值表面 [x,y,Δz] 上的表面均方根粗糙度 s [13],其定義如式(2)所示

式中: A 為取樣面積; Z(x,y) 為所定義區(qū)域內(nèi)各點(diǎn)的高度差值。重復(fù)掃描并評(píng)價(jià)同一位置的臺(tái)階高度平均值
,通過(guò)式(3)來(lái)評(píng)估WLIS的縱向
分辨力

本文使用 8nm 臺(tái)階高度標(biāo)準(zhǔn)樣板對(duì)某WLIS進(jìn)行縱向分辨力測(cè)試,測(cè)試結(jié)果如圖3所示。標(biāo)準(zhǔn)樣板的掃描結(jié)果[圖3(a)]中不同顏色代表不同的表面高度。選取一條高度輪廓,如圖3(b)所示,測(cè)得臺(tái)階高度,驗(yàn)證系統(tǒng)具備亞納米量級(jí)的縱向測(cè)量分辨力。

綜上分析,配合相應(yīng)的測(cè)量平臺(tái)和干涉物鏡等硬件,WLIS的縱向測(cè)量范圍可覆蓋亞納米到數(shù)毫米,乃至 10mm 。
WLIS(非掃描)橫向測(cè)量范圍上限取決于干涉物鏡放大倍率和測(cè)量系統(tǒng)視場(chǎng)大小,后者理論上等于干涉物鏡視場(chǎng)大小和圖像傳感器的靶面尺寸兩者的較小值,而一般圖像傳感器靶面尺寸選型略小于視場(chǎng)尺寸,因此實(shí)際系統(tǒng)單次橫向測(cè)量范圍取決于干涉物鏡放大倍率和圖像傳感器靶面尺寸,其跨度范圍從 120μm 左右(100倍干涉物鏡)到數(shù)毫米(2.5倍干涉物鏡)。結(jié)合可沿 x 軸、 y 軸掃描的運(yùn)動(dòng)平臺(tái),通過(guò)拼接算法可以成倍提高橫向測(cè)量范圍上限。
WLIS橫向測(cè)量范圍下限主要受限于系統(tǒng)光學(xué)分辨率 δo ,由數(shù)值孔徑 NA 與光源中心波長(zhǎng) λ0 決定[見(jiàn)式(4)]。中心波長(zhǎng)越小,數(shù)值孔徑越大,放大倍數(shù)越大,則光學(xué)分辨率越小。一般橫向測(cè)量范圍的下限不小于 0.3μm(100 倍干涉物鏡)。

3.2.3 示值誤差
WLIS的縱向測(cè)量示值誤差、橫向測(cè)量示值誤差及其測(cè)量不確定度可以評(píng)估系統(tǒng)測(cè)量準(zhǔn)確度。縱向測(cè)量示值誤差的影響量如表1所列,其中,光源特性、掃描參數(shù)的設(shè)定、測(cè)量平臺(tái)的穩(wěn)定性以及測(cè)量算法準(zhǔn)確性的影響較顯著。光源決定了光強(qiáng)信號(hào)的一致性與干涉條紋的可見(jiàn)度,直接影響了用于算法分析計(jì)算的原始干涉信號(hào);掃描參數(shù)的設(shè)定與測(cè)量平臺(tái)的穩(wěn)定性決定了測(cè)量數(shù)據(jù)的偏差程度,直接影響了用于算法分析計(jì)算的原始測(cè)量信號(hào);測(cè)量算法是對(duì)原始干涉信號(hào)與原始測(cè)量信號(hào)的處理過(guò)程,其算法準(zhǔn)確性決定了高度位置重構(gòu)的準(zhǔn)確性,即縱向示值的準(zhǔn)確性。一般而言,WLIS的縱向測(cè)量示值誤差可以達(dá)到納米級(jí)至亞納米級(jí)水平。橫向測(cè)量示值誤差的較顯著的影響量主要來(lái)自物鏡的數(shù)值孔徑、圖像傳感器的像素尺寸與放大倍率。
可以使用微納米三維幾何量標(biāo)準(zhǔn)樣板對(duì)WLIS進(jìn)行校準(zhǔn)。微納米三維幾何量標(biāo)準(zhǔn)樣板包含臺(tái)階高度和二維線間隔特征(如圖4所示),用來(lái)確定系統(tǒng)縱向與橫向示值誤差水平。校準(zhǔn)用的樣板的準(zhǔn)確度水平應(yīng)高于WLIS相應(yīng)的標(biāo)稱技術(shù)指標(biāo),如標(biāo)準(zhǔn)樣板的不確定度應(yīng)小于WLIS的最大允許誤差。

WLIS測(cè)量臺(tái)階高度和線間隔的測(cè)量模型與 下述分量有關(guān)[14]


式中: H 、 P 分別為被測(cè)參數(shù)理論測(cè)量值;
)
分別為多次測(cè)量的平均值; δstage 為測(cè)量運(yùn)動(dòng)平臺(tái)引入的測(cè)量誤差; δopt 為光學(xué)系統(tǒng)引入的誤差; δT 為溫度影響引入的誤差; δC 為所用線間隔校準(zhǔn)樣板定值不準(zhǔn)確引入的誤差; θxyz 包括手動(dòng)選取的線間隔截線正交偏差和光學(xué)部件工作面之間夾角引入的誤差。
3.2.4 WLIS的量值溯源
WLIS在微納米幾何量計(jì)量中的溯源鏈如圖5所示,其上一級(jí)應(yīng)為對(duì)其進(jìn)行校準(zhǔn)的標(biāo)準(zhǔn)樣板,這些樣板將繼續(xù)向上溯源,通過(guò)計(jì)量型的微納米幾何量標(biāo)準(zhǔn)裝置進(jìn)行溯源,直至國(guó)際單位制(SI)米的定義[15]
評(píng)價(jià)指標(biāo)和校準(zhǔn)方法,該方法能夠確保WLIS可溯源性。
作為常用的微納米計(jì)量?jī)x器,白光干涉測(cè)量系統(tǒng)的定量測(cè)量能力驗(yàn)證和校準(zhǔn)具有非常實(shí)際的意義。通過(guò)可瀕源的WLIS的校準(zhǔn)方案保障儀器測(cè)量結(jié)果的準(zhǔn)確性、一致性和可靠性,從而保障生產(chǎn)應(yīng)用中相關(guān)產(chǎn)品的質(zhì)量。
參考文獻(xiàn):
[1]DE GROOT P, DECK L. Surface profiling by analysis of white-light interferograms in the spatial frequency domain[J]. Jourmal of Modern Optics,1995,42(2): 389-401.
[2]SCHMIT J, OLSZAK A. High-precision shape measurement by white-light interferometry with realtime scanner error correction[J]. Applied Optics, 2002, 41(28):5943-5950.
[3]DECK L, DE GROOT P. High-speed noncontact profiler based on scanning white-light interferometry[J]. Applied Optics, 1994,33(31): 7334-7338.
[4]蘇榕,劉嘉宇,喬瀟悅,等.用于表面形貌測(cè)量的掃描 白光干涉技術(shù)進(jìn)展[J].激光與光電子學(xué)進(jìn)展,2023, 60(3): 0312005.
[5]WANG S Z, XIE T B, CHANG S P. Vertical scanning white light interfering profilometer based on Linnik interference microscope[C]//Proceedings of SPIE 7656, 5thInternational Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test and Measurement Technology and Equipmen. Dalian: SPIE,2010:765631.
[6]郭彤,胡春光,陳津平,等.垂直掃描白光干涉術(shù)用于 微機(jī)電系統(tǒng)的尺寸表征[J].光學(xué)學(xué)報(bào),2007,27(4): 668 -672.
[7]鄧林涓.白光干涉術(shù)的測(cè)量系統(tǒng)與處理算法的研究 [D].杭州:中國(guó)計(jì)量大學(xué),2015.
[8]郁道銀,談恒英.工程光學(xué)[M].4版.北京:機(jī)械工業(yè) 出版社,2016.
[9]KINO G S,CHIM S S C. Mirau correlation microscope[J]. Applied Optics, 1990, 29(26): 3775- 3783.
[10]中華人民共和國(guó)國(guó)家質(zhì)量監(jiān)督檢驗(yàn)檢疫總局.JJF

4結(jié)論
本文基于對(duì)白光干涉測(cè)量系統(tǒng)原理的研究,按組成模塊分別論述剖析了影響系統(tǒng)測(cè)量結(jié)果的因素;分析了由這些因素決定的WLIS的計(jì)量特性,闡述了國(guó)家標(biāo)準(zhǔn)中計(jì)量特性參數(shù)/曲線與WLIS實(shí)際應(yīng)用時(shí)的表現(xiàn)之間的聯(lián)系;結(jié)合儀器作為工作計(jì)量器具的作用,提出了WLIS的主要
1001—2011通用計(jì)量術(shù)語(yǔ)及定義[S].北京:中國(guó)質(zhì)檢出版社,2012.
[11]國(guó)家市場(chǎng)監(jiān)督管理總局,國(guó)家標(biāo)準(zhǔn)化管理委員會(huì).GB/T33523.604—2022產(chǎn)品幾何技術(shù)規(guī)范(GPS)表面結(jié)構(gòu)區(qū)域法第604部分:非接觸(相干掃描干涉)式儀器的標(biāo)稱特性[S].北京:中國(guó)標(biāo)準(zhǔn)出版社,2022.
[12]WEIJS,SUNKX,CAIXY.Calibration ofWLIlateral indication error with 2D micro/nano pitchstandard[J].Journal ofPhysics:ConferenceSeries,2021,1732(1): 012138.
[13]中華人民共和國(guó)國(guó)家質(zhì)量監(jiān)督檢驗(yàn)檢疫總局,中國(guó)國(guó)家標(biāo)準(zhǔn)化管理委員會(huì).GB/T33523.2—2017產(chǎn)品幾何技術(shù)規(guī)范(GPS)表面結(jié)構(gòu)區(qū)域法第2部分:術(shù)語(yǔ)、定義及表面結(jié)構(gòu)參數(shù)[S].北京:中國(guó)標(biāo)準(zhǔn)出版社,2017.
[14]蔡瀟雨,魏佳斯,孫愷欣,等.白光干涉測(cè)量系統(tǒng)的測(cè)量不確定度評(píng)定[J].計(jì)量學(xué)報(bào),2021,42(6):731-737.
[15] Consultative Committee for Length. CCL strategydocument 2018-2028[R].Paris: BIPM,2018.
(編輯:張磊)