馬興 郭貴勇


摘? 要:該文介紹了彎折儀的工作原理和用途,建立了彎折儀校準結果的測量模型,分析了各不確定度分量主要來源,結合測量實例具體數據,對其合成不確定度進行分析評定,對各不確定度分量進行計算并得出合成不確定度和擴展不確定度,最終可得出該項目的校準測量能力,為用于高分子卷材和防水涂料的低溫彎折性試驗提供了技術支撐。
關鍵詞:彎折儀? 校準方法? 測量? 性能影響? 不確定度評定
Abstract: This paper introduces the working principle and application of the bend machine, establishes the measurement model of the calibration results of the interface tension meter, analyzes the main sources of each uncertainty component, analyzes and evaluates the synthetic uncertainty combined with the concrete data of the measurement example, calculates each uncertainty component, obtains the synthetic uncertainty and extended uncertainty. Finally, the calibration and measurement ability of the project can be obtained. Overall, this paper can provide strong technical support for low temperature bending test of polymer coil and waterproof coating.
彎折儀適合做防水卷材的低溫折彎試驗,又適合防水涂料的低溫柔性測定。彎折儀主要由金屬材料制成的上、下平板,轉軸和調距螺桿組成;平板間距可任意調節。
1? 彎折儀的校準
彎折儀的校準是使用電子數顯示卡尺,測量彎折儀上行板的上端寬度、下端寬度、厚度及上下端距離,上述項目均測量3次取測量值的平均值;測量彎折儀下行板的上端寬度、下端寬度、厚度及上下端距離,上述項目均測量3次取測量值的平均值。
2? 彎折儀的示值誤差不確定評定
2.1 測量模型
2.2 標準不確定度分量的評定
(1)輸入量的標準不確定度u()的評定:其不確定度來源主要是測量重復性引入的不確定度,可以A類方法評定u()。
(2)選取一彎折儀,以電子數顯卡尺對彎折儀上行板的上端寬度連續進行3次測量,得到測量如表1所示。
其所求的平均值:=120.1mm
單次實驗標準偏差:極差法計算(系數C=1.69):s=0.0/1.69mm=0mm。
實際測量中是在重復性條件下測量3次,取其值為測量結果。則=0mm。
(3)電子數顯卡尺的最大允許誤差為±0.03mm,估計為均勻分布,其引入的標準不確定度為:
(4)把u1()與u2()兩項合成,則u()=。
2.3 合成標準不確定度的評定
(1)靈敏系數。
(2)標準不確定度匯總表(見表2)。
(3)合成標準不確定度的計算。
2.4 擴展不確定度的評定
2.5 擴展不確定度的報告與表示
測量彎折儀上行板的上端寬度時,其標準值測量結果擴展不確定度為U=0.1mm。它是由合成標準不確定度uc=0.08mm和包含因子k=2之乘積得到。
2.6 彎折儀各測量點測量結果不確定度評估
根據GB/T 328.15-2016規定,彎折儀上行板的上端寬度、下端寬度、厚度及上下端距離分別為120mm、180mm、10mm、120mm;測量彎折儀下行板的上端寬度、下端寬度、厚度及上下端距離分別為120mm、160mm、17mm、155mm,其測量不確定度如表3所示。
3? 結語
保證彎折儀的準確性,通過校準等方法實現量值的溯源性,在防水涂料和防水卷材等行業的質量控制中起到重要作用。該文介紹研究院對彎折儀進行校準的方法和不確定度評定的方法。
參考文獻
[1] 國家質量監督檢驗檢疫總局.GB 12953-2011,聚氯乙烯(PVC)防水卷材[S].北京:中國標準出版社,2011.
[2] 國家質量監督檢驗檢疫總局.GB/T 16777-2008,建筑防水涂料試驗方法[S].北京:中國標準出版社,2008.
[3] 國家質量監督檢驗檢疫總局.GB/T 328.15-2016,高分子防水卷材低溫彎折性[S].北京:中國標準出版社,2016.